Scan Timing

Scan Delay


Scan Delay

  • If volume, flow rate, and iodine concentration are kept constant, delaying the scan becomes more critical as the:
    • Number of slices per rotation increases (4 to 8 to 16)
    • Scan becomes faster
    • Scan duration shortens
  • It is essential to increase scan delay to image during peak arterial enhancement

Scan delay: how to optimize image quality during arterial enhancement

  • When using shorter scan durations (16-row scanners), consider
    • Decreasing the contrast volume
    • Using contrast solutions with high iodine concentration (the higher, the better)
    • Using a high flow rate (≥4 mL/s) for contrast administration
    • Increasing the scan delay compared with 4-row scanners
    • Using bolus tracking
  • When using longer scan durations (4-row scanners or extended coverage), consider
    • Increasing the contrast volume
    • Using a lower flow rate to increase the injection duration
    • Using contrast solutions with a high iodine concentration to increase the enhancement level for a given injection rate
    • Using bolus tracking or automated settings provided by the scanner manufacture

Scan delay: how to optimize image quality during hepatic parenchymal enhancement

  • Iodine dose is critical
    • Use higher volumes of lower iodine concentration
    • Use lower volumes of higher iodine concentration
    • Iodine dose should not be reduced even if faster scanners are used
  • Flow rate is not as critical as in arterial enhancement
  • Scan delay should be increased if faster scanners are used

  figure7

Representative scan timing and scan duration curves for single-slice, and 4- and 16-slice MDCT scanners are shown for (a) aortic and (b) hepatic enhancement.

As scanners become faster, scan durations become shorter, and therefore it is critical to delay scanning to capture images during peak enhancement, as illustrated in Panel b. Images courtesy of Kyongtae Ty Bae, MD, PhD, Mallinckrodt Institute of Radiology, St. Louis, MO.

 

 

KT Bae

Multidetector CT Protocols
Developed for GE, Philips, Siemens,Toshiba Scanners
Springer 2005, 2006

 
LANGUAGE
Sign In
EMAIL
 
PASSWORD
 
REMEMBER ME ON THIS COMPUTER
FORGOT YOUR PASSWORD?   CLICK HERE
NOT REGISTERED?   CLICK HERE
Calculators
Dose reduction strategies in contrast-enhanced CT:
Radiation Dose Reduction:
A Literature Overview
C. Catalano
Abdominal MDCT
AJ Aschoff
C. Catalano
Read more »
e Book
E-books
 
 
MDCT Anatomy-Body
L. Romano, M. Silva, S. Fulciniti, A. Pinto, Springer 2010
Archive
CT Colonography Bulletin
 
 
CT Colonography Bulletin - Fall 2010
Highlighting a new non invasive technique for colorectal cancer screening
Archive
Services
SpringerLink
Utilities
Bracco - Life from Inside
 
 
Copyright @ Springer 2012

Disclaimer Privacy Terms of Use
Design & Development by Smm
Page Updated May 2008